Materials characterization (SEM, TEM, XRD, XPS)

What is the main reason specimens for TEM (transmission electron microscopy) must be thinned to about 100 nm or less?

  1. AElectrons interact strongly with matter, so they cannot pass through thick specimens, and multiple scattering blurs the image
  2. BThe electron wavelength is longer than that of X-rays, so diffraction does not occur unless the specimen is thin
  3. CThinning the specimen increases charging by the electron beam and raises the contrast
Show answer and explanation

Answer: A. Electrons interact strongly with matter, so they cannot pass through thick specimens, and multiple scattering blurs the image

Because electrons are charged, they interact with matter far more strongly than X-rays, and even at an accelerating voltage of about 200 kV a thick specimen transmits almost nothing or scatters the electrons many times, blurring images and spectra. Specimens are therefore thinned to about 100 nm or less by ion milling or FIB (focused ion beam) for observation. The wavelength of 200 kV electrons is about 0.0025 nm, far shorter than the roughly 0.154 nm of Cu Kα radiation, so the choice saying the wavelength is longer is backward. Charging is a source of image disturbance, not the purpose of thinning. In failure analysis, the standard procedure is to use FIB to thin a lamella targeting the defective site and observe the cross section by TEM.

Report an error

Practice this topic Back to topic

Related calculators

More Materials characterization (SEM, TEM, XRD, XPS) questions

  1. Which correctly describes the information depth of XPS (X-ray photoelectron spectroscopy) and the information it provides?
  2. In XPS using Al Kα radiation (energy 1486.6 eV), the kinetic energy of Si 2p photoelectrons was measured as 1378.8 eV. If the spectrometer work function is 4.5 eV, which value is closest to the binding energy?
  3. In semiconductor failure analysis, which technique is best suited to measuring the depth profile of dopant concentration with sensitivity at the ppm to ppb level?
  4. When a polymer specimen is measured while being heated, which technique is suited to detecting the glass transition temperature Tg, and how does the transition appear?
  5. When calcium oxalate monohydrate (CaC₂O₄·H₂O, molar mass 146.1 g/mol) is heated in TGA, the water of crystallization is released first. Which value is closest to the percent mass loss in this first step? Take the molar mass of H₂O as 18.0 g/mol.